Test object for easy qualification of computed radiography systems according to the latest standards (ISO 16371-1, ASTM E 2445)
Ultra-compact
With a height of only 20 cm and a width of 27.5 cm, it is the smallest CR Phantom on the market.
The CR Phantom can test all important performance characteristics of a CR scanner system, including basic spatial resolution (unsharpness), contrast, MTF, laser beam jitter, scanner slip and scanner shading. These tests are described in detail in standards ISO 16371-1 and ASTM E 2445 and should be performed regularly to ensure correct and accurate system operation. For easy use of the CR Phantom, a test procedure and D-Tect Viewer software are available as a free download.
To meet the requirements of these standards, the DÜRR NDT CR Phantom includes two duplex wire IQIs and measurement points for shading correction positioned in both axis directions (horizontal and vertical). This makes it possible to map all required information on the imaging plate with a single X-ray exposure – the CR Phantom does not need to be rotated to obtain data from the other axis, resulting in more precise test results and significant time savings.
Ready for use
Including report template and test procedure
T-target (brass)
Laser beam jitter, MTF check, blooming (flaring)
Duplex wire IQI 15D (high resolution)
Basic spatial resolution (unsharpness)
BAM snail
Central beam alignment
Line pair IQI type 53-b
Line pair resolution
Measurement points
Shading correction

Well protected
A padded wooden case protects the CR Phantom during storage and transport
Cassette position locator
Cassette (imaging plate) positioning
Homogeneous strip (aluminium)
Scanner slip, scanner shading
Lucite plate
Support plate
cm/inch ruler
Linearity check
Contrast sensitivity gauge
Contrast sensitivity check